| 3.4.1 Stop-Start Problem | ||
| 3.4.2 No Correlation |
When CAOBS is first started, it is possible for the correlator and CAOBS to be out of sync. In this case, there will be no update of data in SPD or VIS and the CACOR window will have a new blue-coloured message every cycle reporting something like:
DELBAT...
The solution is to issue a stop in CAOBS, then restart observing.
There are a number of problems that may result in loss of correlation.
| 3.4.2.1 Conversion System Out of Lock |
It is possible to lose lock in the conversion system. Check the following in MONICA. Currently there isn’t a standard page with these in it and you have to create your own pages in MONICA (see below).
To create a new page in MONICA that monitors the CACOR conversion system
In MONICA choose
Window → New Window → Setup → Add panel → Time Series
This brings up the page for setting up a time series.
Optical Power Photodiode biases need to be a reasonable value (if < 0.2mA will lead to loss of lock)
Won’t get any sync errors, correlator will look ok, but no correlation because tuned to the wrong frequency.