Next: Local Oscillator Software
Up: VIS
Previous: VIS Commands
Contents
Index
CACAL
CACAL is a calibration tool that you should use at the start of
observations. You do not need to run CACAL again unless:
- you change frequencies;
- a major problem occurs and the delays differ from zero by more
than 2ns for continuum observations or 20ns for spectral line
observations or 0.5ns for 1934-638.
Before running CACAL, check that the delays are flat.
CACAL calibrates the delays, amplitude and phase for each
baseline. The latter two calibrations are not essential, as the
offline calibration can deal with them. To use CACAL, perform the
following steps:
- Ensure that the Tsys field on CAOBS reads ON. (If not enter
CAOBS
enable tsys).
- Wait until the array is on source for a strong calibrator (like
1934-638), and the attenuators have adjusted.
- If you have bad interference, use the CACOR command line to select
the channels to be loaded as the visibility variables for CACAL and VIS
to use. Use tvchans
- Ensure that delays are stable on all baselines and all polarisations.
- Locate a terminal logged on to XBONES.
- Enter the command XBONES
cacal.
- Answer A to Automatic or [I] Interactive.
- Check the results with VIS - note that it takes a few cycles
for the data to be displayed.
After running CACAL, check that the delays, amplitudes and phases
all seem reasonable.
Strong interference can produce a poor
calibration. If there is significant interference, you
may want to run CACAL on a smaller number of correlator channels.
Next: Local Oscillator Software
Up: VIS
Previous: VIS Commands
Contents
Index
Robin Wark
2006-10-24