Starting Observations
At the start of observations, the observer needs to load the schedule file, ensure that sampler settings are in range and calibrate the array.
The Duty Astronomer should be available to help observers starting their observations
Procedure
Load the setup schedule file | The schedule file for calibration needs to be loaded into caobs caobs> set file schedule_file | ||||||||||
Observe the calibrator | In caobs, drive to the calibrator with caobs> track # (where # is the
scan number within the schedule file) | ||||||||||
Set sampler levels | Once on source, caobs should attempt to set attenuators so that the power levels into the samplers are within the sampler range. If there is insufficient range with the fine attenuators the observer may need to set coarse attenuators and occasionally even mm attenuators. | ||||||||||
Check visibilities | Once the attenuators are stable (no automatic changes for more than 1 minute)
visibilities need to be checked to ensure that they are stable. In vis, check the delays, amplitudes and phases are stable for all baselines. | ||||||||||
cacal | If visibilities are stable, continue tracking on the calibrator, and in xbones
window, run cacal with xbones> cacal In general, it is appropriate to accept automatic mode. | ||||||||||
amplitude calibration | At some point, amplitude and bandpass calibration data needs to be observed. For cm observations (including 1cm observations) this will (in general) be a short (5-10 min) observation on PKS1934-638. 3mm calibration is more complicated: See XXXX for information on 3mm calibration If the calibrator is above 30 degrees at the start of the observation, it is a good time to do the calibration. If calibration cannot be done at the start, time has to be allowed at the end of the observation. To observe the amplitude calibrator:
| ||||||||||
start observing | Finally, real observing can begin
|
Original: Robin Wark (17-Aug-2006)
Modified: Robin Wark (24-Aug-2006)